Validation & System Suitability
Check system-suitability results against the harmonised pharmacopoeial criteria and see what a full validation requires.
System-suitability check
USP ⟨621⟩, harmonised text: symmetry factor 0.8–1.8 (official 1 May 2025). Maximum permitted RSD from the harmonised table for the assay of a pure substance.
3 to 6
e.g. limit 102.0% → B = 2.0
from the monograph or your method
| Parameter | Value | Criterion | Result |
|---|---|---|---|
| System repeatability (RSD %) | 0.6 | ≤ 0.85% (table: n = 6, B = 2%) | PASS |
| Symmetry factor (As) | 1.2 | 0.8–1.8 | PASS |
| Resolution (Rs) | 2.5 | ≥ 2 (method / monograph) | PASS |
Column-performance recommendations
Not pharmacopoeial requirements. Commonly cited recommendations, from FDA’s 1994 reviewer guidance on chromatographic methods; they do not change the verdict above.
| Theoretical plates (N) | 6000 | > 2000 (recommended) | OK |
| Retention factor (k) | 4 | > 2 (recommended) | OK |
A monograph or an approved method can set different limits; those always take precedence. Always confirm against the current pharmacopoeia.
Validation requirements planner
| Validation parameter | Typical industry acceptance criterion |
|---|---|
| Specificity | No interference at the analyte retention time |
| Linearity | r ≥ 0.999 over 80–120% of nominal |
| Accuracy | Recovery 98–102% (mean) |
| Repeatability | RSD ≤ 2% (n ≥ 6) |
| Intermediate precision | RSD ≤ 2% across days/analysts |
| Range | 80–120% of test concentration |
| Robustness | Deliberate small changes — results within criteria |
ICH Q2(R2) and USP ⟨1225⟩ define which parameters to validate for each type of procedure but do not set numeric limits; the assay, impurity and dissolution figures above are common industry practice. The bioanalytical figures are those stated in ICH M10. Always confirm against your product specification, your quality system and the current pharmacopoeia.